Home
»
Honeywell Midas® S2 Sensor Cartridges
»
Midas® S2 Sensor Cartridges Bromine (Br2), Chlorine Dioxide (ClO2)0.036-0.4 ppm P/N: MIDAS S2-E-BR2
Midas® S2 Sensor Cartridges Bromine (Br2), Chlorine Dioxide (ClO2)0.036-0.4 ppm P/N: MIDAS S2-E-BR2
Regular price
£331.00 GBP
Contact ISS
Midas® S2 Sensor Cartridges Bromine (Br2), Chlorine Dioxide (ClO2)0.036-0.4 ppm P/N: MIDAS S2-E-BR2
| Selectable gases |
Chlorine (Cl2) Fluorine (Br2) Chlorine Dioxide (ClO2) |
| Cartridge Part Number | MIDAS S2-E-BR2 2-year extended warranty |
| Sensor Technology | Electrochemical sensor |
| Measuring Range | Cl2 0 – 0.4ppm Br2 0 – 0.4ppm ClO2 0 – 0.4ppm |
| Default Alarm 1 | Cl2 0.05ppm Br2 0.05ppm ClO2 0.05ppm |
| Default Alarm 2 | Cl2 0.1ppm Br2 0.1ppm ClO2 0.1ppm |
| LDL, LAL | Cl2 0.036ppm Br2 0.036ppm ClO2 0.036ppm |
| Resolution | Cl2 0.002ppm Br2 0.002ppm ClO2 0.002ppm |
| Accuracy | ≤ ± 5% of measured value |
| Response Time t62.5 | Typical 2 seconds |
| Sensor Cartridge Life Expectancy (Expiration Period) |
24 months under typical application conditions Extendable for 1 year through calibration after 24 months |
| Operating Temperature Effect of Temperature Sensitivity |
0°C to +40°C (32°F to 104°F) ≤ ± 10% of measured value at 20°C |
| Operating Humidity | 10 to 90% non-condensing |
| Operating Pressure | 90 – 110kPa |
| Calibration Gas | Cl2 0.2ppm Br2 0.2ppm ClO2 0.2ppm |
| Warm Up Time | < 10 minutes |
| Storage Temperature | +5°C to +25°C (+41°F to +77°F) |