Dräger Tube -Triethylamine 5/a
Contact ISS

Dräger Tubes For Short Term Measurements: Triethylamine 5/a
Dräger Safety offer a massive range of short-term gas detection tubes for the measurement of various gases. This particular set of tubes is for (Triethylamine 5/a).
The tubes come in a pack of 10 tubes and have a shelf life of 2 years.
| Part no | Dräger Tube | Substance | Measurement Range |
| 6718401 | Triethylamine 5/a | n-Butylamine | 5 to 60 ppm |
| Diethylamine | 5 to 60 ppm | ||
| Diisopropylamine | Qualitative | ||
| Dimethylethylamine | 5 to 60 ppm | ||
| Dimethylisopropylamine | 15 to 70 ppm | ||
| Dipropylamine | Qualitative | ||
| Ethylamine | 5 to 60 ppm | ||
| Isopropyl Amine | 5 to 60 ppm | ||
| Methylamine | 5 to 60 ppm | ||
| N-Methyldiethylamine | 5 to 60 ppm | ||
| n-Propylamine | 5 to 60 ppm | ||
| iso-Propylamine | 5 to 60 ppm | ||
| Tributylamine | 2.5 to 50 ppm | ||
| Triethylamine | 5 to 60 ppm | ||
| Trimethylamine | 5 to 60 ppm |
Application Range:
| Standard Measuring Range: | 5 to 60 ppm |
| Number of Strokes n: | 5 |
| Time for Measurement: | approx. 3 min |
| Standard Deviation: | ± 10 to 15 % |
| Colour Change: | yellow to blue |
Ambient Operating Conditions:
| Temperature: | 10 to 40 °C |
| Absolute Humidity | 5 to 12 mg H2O / L |
Reaction Principle:
(C2H5)3N + acid ~> blue reaction product
Cross Sensitivity:
Other basic substances such as organic amines and ammonia are indicated, but with different sensitivities..