Description
Dräger Tubes For Short Term Measurements: Dimethyl Sulfide 1/a
Dräger Safety offer a massive range of short-term gas detection tubes for the measurement of various gases. This particular set of tubes is for (Dimethyl Sulfide 1/a).
The tubes come in a pack of 5 tubes and have a shelf life of 2 years.
Part no | Dräger Tube | Substance | Measurement Range |
6728451 | Dimethyl Sulphide 1/a | Acetaldehyde | Quantitative |
Acrolein | 0.1 to 10 ppm | ||
Dimethyl Sulphide | 1 to 60 ppm |
Application Range:
Standard Measuring Range: | 1 to 15 ppm |
Number of Strokes n: | 20 |
Time for Measurement: | approx. 15 min |
Standard Deviation: | ± 15 to 30 % |
Colour Change: | violet to yellow brown |
Ambient Operating Conditions:
Temperature: | 15 to 30 °C |
Absolute Humidity | < 20 mg H2O / L |
Reaction Principle:
(CH3) 2S + KMnO4 ~> MnIV + various oxidation products
Cross Sensitivity:
Many organic compounds with C=C double bonds are indicated as well, but with different sensitivities. It is impossible to differentiate between them. H2S (hydrogen sulfide) is indicated with approximately twice the sensitivity. The tube H2S 5/b can be used as a filtering tube. Then, with n = 20 pump strokes approx. 30 ppm H2S is retained. Methyl mercaptan is indicated with approximately twice the sensitivity..
For more information on Dräger-Tubes click the image below.
To view our range of Dräger pumps please click below.