Dräger Tube -Carbon Dioxide 0.1%/a

DrägerSKU: CH23501

Price:
Sale price£62.89

Description

Dräger Tubes For Short Term Measurements: Carbon Dioxide 0.1%/a

Dräger Safety offer a massive range of short-term gas detection tubes for the measurement of various gases. This particular set of tubes is for (Carbon Dioxide 0.1%/a).

The tubes come in a pack of 10 tubes and have a shelf life of 2 years.

Application Range:

Standard Measuring Range: 0.5 to 6 vol. % / 0.1 to 1.2 vol. %
Number of Strokes n: 1
Time for Measurement: approx. 30 s / approx. 2.5 min
Standard Deviation: ± 5 to 10 %
Colour Change: white to violet

Ambient Operating Conditions:

Temperature: 0 to 30 °C
Absolute Humidity max. 30 mg H2O /L

Reaction Principle:

CO2 + Amine ~> violet reaction product

Cross Sensitivity:

No influence on the reading by 10 ppm hydrogen sulfide and 2 ppmsulfur dioxide..

For more information on Dräger-Tubes click the image below.

To view our range of Dräger pumps please click below.