Dräger Tube -Aniline 0.5/a

DrägerSKU: 6733171

Price:
Sale price£105.72

Description

Dräger Tubes For Short Term Measurements: Aniline (0.5/a)

Dräger Safety offer a massive range of short-term gas detection tubes for the measurement of various gases. This particular set of tubes is for Aniline (0.5/a)

The tubes come in a pack of 5 tubes and have a shelf life of 2 years.

Part no Dräger Tube Substance Measurement Range
6733171 Aniline 0.5/a Aniline 0.5 to 10 ppm
N,N-Dimethylaniline 0.5 to 10 ppm

Application Range:

Standard Measuring Range: 0.5 to 10 ppm
Number of Strokes n: 20
Time for Measurement: approx. 4 min.
Standard Deviation: ± 15 to 20 %
Colour Change: pale yellow to pale green.

Ambient Operating Conditions:

Temperature: 15 to 30°C
Absolute Humidity 7 to 12 mg H2O/L

Reaction Principle:

C6H5NH2 + CrVI ~> CrIII + various oxidation products

Cross Sensitivity:

When methyl aniline and aniline are present at the same time, it is impossible to measure only the aniline concentration. Ethers, ketones, esters, aromatics and benzines are indicated as well but with different sensitivities.