Dräger Tube -Acetone 40/a

DrägerSKU: 8103381

Price:
Sale price£65.76

Description

Dräger Tubes For Short Term Measurements: Acetone (40/a)

Dräger Safety offer a massive range of short-term gas detection tubes for the measurement of various gases. This particular set of tubes is for Acetone (40/a).

The tubes come in a pack of 10 tubes and have a shelf life of 2 years.

Application Range:

Standard Measuring Range: 40 to 80 ppm
Number of Strokes n: 1
Time for Measurement: approx. 1 min.
Standard Deviation: ± 15 to 20 %
Colour Change: pale yellow to yellow

Ambient Operating Conditions:

Temperature: 5 to 40°C
Absolute Humidity 5 to 40 mg H2O/L

Reaction Principle:

Acetone + 2.4-Dinitrophenylhydrazine ~> yellow hydrazone.

Cross Sensitivity:

Other ketones are indicated but with different sensitivities. Aldehydes are indicated. 500 ppm ethylacetate does not affect the indication. Ammonia causes the indicating layer to turn yellowish brown.

For more information on Dräger-Tubes click the image below.

To view our range of Dräger pumps please click below.

You may also like

Recently viewed