Menu
Cart 0
Dräger Tube -Acetone 100/b

Dräger Tube -Acetone 100/b

CH22901
  • £58.97


Dräger Tubes For Short Term Measurements: Acetone (100/b)

Dräger Safety offer a massive range of short-term gas detection tubes for the measurement of various gases. This particular set of tubes is for Acetone (100/b).

The tubes come in a pack of 10 tubes and have a shelf life of 2 years.

Part no Dräger Tube Substance Measurement Range
CH22901 Acetone 100/b Acetone 100 to 12000 ppm
  2-Butanone 100 to 12000 ppm
  2-Heptanone Qualitative
  Hexone (MIBK) 130 to 15600 ppm
  Methyl-n-Amyl Ketone (2-Heptanone) Qualitative
  Methyl Ethyl Ketone 100 to 12000 ppm
  Methyl Isobutylketone (4-Methyl-2-pentanone) 130 to 15600 ppm
  Methylpropyl Ketone 130 to 15600 ppm
    2-Pentanone 130 to 15600 ppm

 

Application Range:

Standard Measuring Range: 100 to 12,000 ppm
Number of Strokes n:  10
Time for Measurement: approx. 4 min.
Standard Deviation: ± 15 to 20 %
Colour Change: pale yellow to yellow

 

Ambient Operating Conditions:

Temperature: 0 to 40°C
Absolute Humidity 5 to 20 mg H2O/L

 

Reaction Principle:

Acetone + 2.4-Dinitrophenylhydrazine ~> yellow hydrazone.

Cross Sensitivity:

Other ketones are indicated but with different sensitivities. Aldehydes are indicated, but not esters. Ammonia causes the indicating layer to turn yellowish brown.

For more information on Dräger-Tubes click the image below.

To view our range of Dräger pumps please click below.


We Also Recommend